3 Removing and fitting housing
3–28 Section 3 © Siemens VDO Trading GmbH TU00-1318-0200302
3.8.3 Removing stylus unit
Fig. 3-19 Removing stylus unit
Procedure
Unhook the tension spring (4) (0.25 x 3 x 17) on the back of the frame (2) or
tension spring (4.1) (0.25 x 3 x 15.3).
Unhook the tension spring (6).
Lift the guide rod (5) out of the frame bracket (1).
Lift guide rod (3) out of the bracket(1).
Lift out the complete stylus unit.
When doing this, mind the anti-twist devices (8).
Remove guide rods (3) and (5).
Replace the stylus carriage.
1 Frame bracket 5 Guide rod, long
2 Frame 6 Tension spring for v carriage
3 Guide rod, short 7 Contact path
4 Tension spring for distance carria-
ge
8 Anti-twist device
4.1 Tension spring for distance carria-
ge
8
8
8
6
5 4.1
from
4
to 08.91
3
5
to 04.90
7
Remark
• The respective stylus holder is firmly bonded to the carriage.
• Should the bonding be loose replace the whole stylus carriage.
• Do not touch the contact path (7). If you do clean and polish it.