Appendix C: Troubleshooting
Wrong Reading
If the Reading is erratic, or jumps to some incorrect value periodically, ensure the:
1. Surface monitored is not beyond the HydroRanger 200’s programmed range or the
transducer’s maximum range.
2. Material is not falling into the transducer’s acoustic beam.
3. Material is not inside the blanking distance of the transducer.
Types of Wrong Readings
If a periodic wrong Reading is always the same value, see Fixed Reading.
If the wrong Reading is random, ensure the material surface to transducer distance is
less than the Empty value entered plus 20%. If the material/object monitored is outside
this distance, increase Range Extension (P801) as required. This error is most common in
OCM applications using weirs.
Liquid Splashing
If the material monitored is a liquid, check for splashing in the vessel. Enter a lower
Maximum Process Speed (P003) value to stabilize the Reading, or install a stilling well.
(Contact Siemens Milltronics or your local distributor.)
Adjust the Echo Algorithm
Use Dolphin Plus to view live sonic profiles and make adjustments to the P820 Algorithm
parameter. See the Parameter Reference for details.
If you do not own Dolphin Plus, connect an oscilloscope and use the hand programmer to
adjust the same parameter.
If the “Area” algorithm is used and narrow noise spikes are evident on the (long shot)
Echo Profile, turn the Spike Filter (P821) on and/or widen the Narrow Echo Filter (P822).
Also, if the true echo has jagged peaks, use Reform Echo (P823).
If multiple echoes appear on the Echo Profile, typical of a flat material profile (especially if
the vessel top is domed), use the “first” Algorithm.
If the Echo Profile repeatedly switches from short to long, adjust the Short Shot Range
(P852) to stabilize the “shot” mode used for the echo evaluation. Also, adjust the Short
Shot Bias to increase (or decrease the amount of preference given to short shot echoes
over long shot echoes.
Should a stable measurement still not be attainable, contact Milltronics or your local
distributor.
Transducer Ringing
If the transducer is mounted too tightly, or if it is mounted so that its side touches
something, its resonance characteristics change and this can cause problems.