Page 146 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01
mmmmm
Parameters
P275 Scaled Discrete Input Value
Shows the current value of the discrete input after any scaling is applied.
Readings are updated continuously even in PROGRAM mode. The value signals a level
override event.
Standard Data Logging (P300 to P321)
All records can be reset by pressing the CLEAR keys.
Record Temperatures (P300 to P303)
These features display the high and/or low temperatures in °C.
When a parameter relating to a TS-3 Temperature Sensor is accessed, the Point Type
display changes to the TS-3 symbol .
If the unit is powered up without a temperature sensor connected, the value –50°C is
displayed. This information can help trace problems with both built in and external
temperature sensors.
Primary Index Discrete input
Values
Display: view only
Values: dependent on the function of the discrete input
Range of Values Function (P270)
1Forced ON
0Forced OFF
0 (DI open), 1 (DI closed) Normally Open
0 (DI closed), 1 (DI open) Normally Closed