Page 198 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01
mmmmm
Parameters
P832 TVT Shaper Adjust
Allows manual adjustment of the TVT curve.
Use this feature to bias the shape of the TVT curve to avoid selecting false echoes from
fixed objects.
Adjustment to this parameter is best done while viewing the echo profile with Dolphin
Plus. Refer to the Dolphin Plus online help for details. If Dolphin Plus is not available, then
an oscilloscope can be used. When using an oscilloscope, the Echo Lock Window display
becomes the TVT Curve Pointer. See
P810 Scope Displays
on page 191 for more
information.
The TVT curve is divided into 40 breakpoints, accessible by enabling the point number as
the breakpoint index field. Each breakpoint is normalized to a value of 0, as displayed in
the parameter value field. By changing the breakpoint value, up or down, the intensity of
the bias applied to that breakpoint of the curve is respectively changed. By changing the
value of adjacent breakpoints, the effective bias to the shaper can be broadened to suit
the desired correction. In the case of multiple false echoes, shaping can be applied along
different points of the curve. Shaping should be applied sparingly in order to avoid
missing the true echo.
Changing a breakpoint:
1. Confirm that P831, TVT shaper, is on.
2. Go to P832.
3. Press PROGRAM twice to highlight the index value.
4. Press ARROWS to scroll through the 40 points (or type in the desired point).
5. Enter the value from –50 to 50.
6. Press .
7. Press TRANSDUCER to update the scope display with the new TVT.
Primary Index Global
Values
Range: -50 to 50
Preset: 0
Related
• P810 Scope Displays
• P831 TVT Shaper