7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 193
mmmmm
Parameters
P816 Echo Time Raw
The time (in ms) from the transmit pulse to the processed echo.
Profile Pointer (P817 to P825)
When one of these parameters is accessed, the scope display changes: The Echo Lock
Window becomes the Profile Pointer. The Profile Pointer may be moved to any point on
the Echo Profile to gain specific information about the profile according to the selected
parameter.
To move the Profile Pointer to a specific point, enter the desired value and it will move to
the nearest acceptable Echo Profile point.
To scroll the Profile Pointer along the Echo Profile:
1. Press to display the Auxiliary Function symbol.
2. Press or to move the Profile Pointer to the left or right respectively.
3. When the Profile Pointer Parameters are exited, the scope display automatically
changes back to include the Echo Lock Window.
P817 Profile Pointer Time
The time (in ms) from the transmit pulse to the Profile Pointer.
Primary Index Transducer
Values Range: 0.0 to 9999 (view only)
Related P815 Echo Time Filtered
Primary Index Global
Values Range: 0.000 to 9999 (view only)
Related
P818 Profile Pointer Distance
P819 Profile Pointer Amplitude
P820 Algorithm
P821 Spike Filter
P822 Narrow Echo Filter
P823 Reform Echo
P825 Echo Marker
Trigger