Parameters
Temperature Compensation (P660 to P664)
P660 Temp Source
Source of the temperature reading used to adjust the speed of sound.
The HydroRanger 200 uses the TS-3 temperature sensor assigned to the transducer
unless one is not connected, in which case the ultrasonic/temperature transducer is
used. If the transducer does not have an internal temperature sensor, the Temp Fixed
(P661) value is used.
If the acoustic beam atmosphere temperature varies with distance from the transducer,
connect a TS-3 Temperature Sensor and ultrasonic / temperature transducers, and select
average.
In gasses other than air, the temperature variation may not correspond with the speed of
sound variation. In these cases turn off the temperature sensor and use a fixed
temperature.
P661 Temp Fixed
Use this feature if a temperature sensing device is not used.
Enter the temperature (in °C) of the atmosphere within the transducer acoustic beam. If
the temperature varies with distance from the transducer, enter the average temperature.
Primary Index Transducer
Values
1*AUTO
2Temp Fixed
3 Ultrasonic/Temperature Transducer
4 TS-3 Temperature Sensor
5 Average (TS-3 and transducer)
Alters • P664 Temperature
Related
• P651 Sound Velocity
• P653 Velocity
• P654 Velocity at 20°C
• P661 Temp Fixed
Primary Index Transducer
Values Range: -199 to 199 (preset = 20 °C)
Related
• P651 Sound Velocity Calibration
• P653 Velocity
• P654 Velocity at 20°C
• P660 Temp Source