7ML19981FC01 HydroRanger 200 – INSTRUCTION MANUAL Page 153
mmmmm
Parameters
Recording a profile manually
Press TRANSDUCER to fire the transducer and record the echo profile into the
internal scope buffer for display.
For differential or average operation (P001 = 4 or 5), access scope Displays (P810)
parameter to select the transducer number.
Saving a manual record
Press ENTER to copy the echo profile record in the scope buffer and save it in the
selected address in the record library. The parameter value field displays the new record
information.
Displaying a record
Press key to enter display auxiliary mode and then:
• Press TRANSDUCER to copy the current echo profile into the scope buffer for
display on an oscilloscope or Dolphin Plus
Deleting a record
Press CLEAR and then ENTER to delete the echo profile record in the selected
address. The value returns to - - - -.
P331 Auto Record Enable
Use to enable/disable the Auto Profile Record function.
Primary Index Global
Values
Range: 0 to 1
0* Off
1On