Siemens HydroRanger 200 Manual de Instruções Página 182

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 181
Page 168 HydroRanger 200 – INSTRUCTION MANUAL 7ML19981FC01
mmmmm
Parameters
P641 Flow Sampler Mantissa
Use this feature in conjunction with Flow Sampler Exponent (P642) to establish the
number of flow units required to increment the Flow Sampler (device connected to the
HydroRanger 200 relay set for the flow sampler operation) (Relay Function, P111 = 41).
This parameter is relevant only if Operation is set to OCM (P001 = 6).
Enter the mantissa (Y) for the exponent (Z) in the formula:
Flow Sampler Increment = Y x 10
Z
Flow units.
Example: To count once every 4310 (4.31 x 10
3
) flow units:
set P641 to 4.31 and P642 to 3
P642 Flow Sampler Exponent
Use this feature in conjunction with Flow Sampler Mantissa (P641) to establish the
number of flow units required to increment the Flow Sampler (device connected to the
HydroRanger 200 relay set for the flow sampler operation Relay Function, P111 = 41).
This parameter is relevant only if Operation is set to OCM (P001 = 6).
Enter the exponent (Z) for the mantissa (Y) in the formula:
Flow Sampler Increment = Y x 10
Z
Flow units.
Primary Index Global
Values
Range: 0.001 to 9999
Preset = 1.000
Related
•P001 Operation
P111 Relay Control Function
OCM (P600 to P621)
P642 Flow Sampler Exponent
Primary Index Global
Values
Range: -3 to +7 (integers only)
Preset = 0
Related
•P001 Operation
P111 Relay Control Function
OCM (P600 to P621)
P641 Flow Sampler Mantissa