See also
Measured value display
(Page 121)
6.5.13 Unit
Introduction
In mode 14, select the physical unit in which the measured value display of the device should
be shown.
Condition
You are familiar with the correct operation of the transmitter and the associated safety notes.
You have already selected the desired measured value display via HART.
Procedure
To adjust the physical unit, proceed as follows:
1. Set mode 14.
2. Select a unit.
Press the [↑] and [↓] buttons simultaneously to set the unit to the first value in the following
table depending on the measuring mode set.
3. Save with the [M] button.
Instructions for selecting the unit
● Unit selection depends on the type of measurement set. For example, only pressure units
are available in "Pressure" measuring mode and only level units are available in the "Level"
measuring mode.
● The displayed measured value is always converted into the new unit. "9.9.9.9.9" appears
on the display when its display capacity is exceeded.
● The selected unit is visible on the display in the measuring mode only if you have selected
the display of a physical unit via HART. "mA" or "%" will be displayed if you have not selected
mode 13, "Measuring mode".
Operation
6.5 Local operation
SITRANS P, DS III series with HART
Operating Instructions, 09/2012, A5E00047092-08 95