Siemens SITRANS P Instruções de Operação Página 31

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 234
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 30
The change in the resistance causes a bridge output voltage proportional to the absolute
pressure.
3.6.3.5
Measuring cell for absolute pressure from the gauge pressure series

S
H
Measuring cell Filling liquid
Process connection Absolute pressure sensor
Seal diaphragm Inlet pressure
Figure 3-11 Function chart of measuring cell for absolute pressure
The inlet pressure (p
e
) is transferred to the absolute pressure sensor via the seal diaphragm
and the fill fluid , displacing its measuring diaphragm. The displacement changes the
resistance of the four piezoresistors (bridge circuit) of the absolute pressure sensor. The
change in the resistance causes a bridge output voltage proportional to the inlet pressure.
3.6.3.6 Measuring cell for gauge pressure, front-flush membrane
S
H
Reference pressure opening Filling liquid
Measuring cell Gauge pressure sensor
Process connection p
e
Inlet pressure
Seal diaphragm
Figure 3-12 Function chart of the measuring cell for gauge pressure, flush mounted diaphragm
Description
3.6 Functional principle
SITRANS P, DS III series with HART
Operating Instructions, 09/2012, A5E00047092-08 29
Vista de página 30
1 2 ... 26 27 28 29 30 31 32 33 34 35 36 ... 233 234

Comentários a estes Manuais

Sem comentários