Rated conditions for absolute pressure (from the differential pressure series), differential pressure and flow rate
● Degree of protection in
accordance with EN 60529
IP65, IP68
● Degree of protection in
accordance with NEMA 250
NEMA 4X
● Electromagnetic compatibility
Emitted interference and
interference immunity
In accordance with EN 61326 and NAMUR NE 21
Process medium conditions
● Process medium temperature
Measuring cell with silicon oil
filling
-40 … +100 °C (-40 … +212 °F)
● Measuring cell
30 bar (435 psi)
● -40 … +85 °C (-40 … +185 °F)
●
With flow: -20 … +85 °C (-4 … +185 °F)
Measuring cell with inert liquid
filling
-20 … +100 °C (-4 … +212 °F)
● Measuring cell
30 bar (435 psi)
● -40 … +85 °C (-40 … +185 °F)
●
With flow: -20 … +85 °C (-4 … +185 °F)
In conjunction with dust
explosion protection
-20 … +60°C (-4 … +140°F)
Rated conditions for level
Installation conditions
● Installation instruction specified through the flange
Ambient conditions
● Ambient temperature
Note Observe the allocation of the max. permissible operating temperature to the max.
permissible operating pressure of the relevant flange connection.
Measuring cell with silicon oil
filling
-40 … +85 °C (-40 … +185 °F)
Display -30 … +85 °C (-22 … +185 °F)
Storage temperature -50 … +85 °C (-58 … +185 °F)
● Climate class
Condensation Permitted
● Degree of protection in
accordance with EN 60529
IP65
● Degree of protection in
accordance with NEMA 250
NEMA 4X
● Electromagnetic compatibility
Emitted interference and
interference immunity
In accordance with EN 61326 and NAMUR NE 21
Medium conditions
● Process temperature
Measuring cell with silicon oil
filling
● Plus side: see the mounting flange
●
Minus side: -40 … +100°C (-40 … +212°F)
Technical data
11.5 Operating conditions
SITRANS P, DS III series with HART
186 Operating Instructions, 09/2012, A5E00047092-08