11.2 Input point
Gauge pressure input
HART PROFIBUS PA or Foundation Fieldbus
Measured variable Gauge pressure
Span (continuously
adjustable) or measuring
range, max. operating
pressure (in accordance
with 97/23/EC Pressure
Equipment Directive) and
max. test pressure (in
accordance with
DIN
16086) (max. 120 bar
for oxygen measurement)
Measuring
span
Maximum
operating
pressure
MAWP (PS)
Maximum
permissible
test pressure
Measuring
range
Maximum
permissible
operating
pressure
Maximum test
pressure
0.01 …
1 bar g (0.15
… 14.5 psi g)
4 bar g
(58 psi g)
6 bar g
(87 psi g)
1 bar g
(14.5 psi g)
4 bar g
(58 psi g)
6 bar g
(87 psi g)
0.04 …
4 bar g (0.58
… 58 psi g)
7 bar g
(102 psi g)
10 bar g
(145 psi g)
4 bar g
(58 psi g)
7 bar g
(102 psi g)
10 bar g
(145 psi g)
0.16 …
16 bar g (2.3
… 232 psi g)
21 bar g
(305 psi g)
32 bar g
(464 psi g)
16 bar g
(232 psi g)
21 bar g
(305 psi g)
32 bar g
(464 psi g)
0.63 … 63 bar
g (9.1 …
914 psi g)
67 bar g
(972 psi g)
100 bar g
(1450 psi g)
63 bar g
(914 psi g)
67 bar g
(972 psi g)
100 bar g
(1450 psi g)
1.6 …
160 bar g (23
… 2321 psi g)
167 bar g
(2422 psi g)
250 bar g
(3626 psi g)
160 bar g
(2321 psi g)
167 bar g
(2422 psi g)
250 bar g
(3626 psi g)
4 … 400 bar g
(58 …
5802 psi g)
400 bar g
(5802 psi g)
600 bar g
(8702 psi g)
400 bar g
(5802 psi g)
400 bar g
(5802 psi g)
600 bar g
(8702 psi g)
7,0 ...
700 bar g
(102 ...
10153 psi g)
800 bar g
(11603 psi g)
800 bar g
(11603 psi g)
700 bar g
(10153 psi g)
800 bar g
(11603 psi g)
800 bar g
(11603 psi g)
Lower measuring limit
● Measuring cell with
silicon oil filling
30 mbar a (0.44 psi a)
● Measuring cell with
inert liquid
30 mbar a (0.44 psi a)
Upper measuring limit 100% of max. span (max. 120 bar g
(1740 psi g) for oxygen measurement)
100% of max. measuring range (max. 120 bar g
(1740 psi g) for oxygen measurement)
Start-of-scale value between the measuring limits (continuously adjustable)
Technical data
11.2 Input point
SITRANS P, DS III series with HART
170 Operating Instructions, 09/2012, A5E00047092-08
Comentários a estes Manuais