Technical specifications (continued)
Industrial Wireless Communication
WirelessHART
SITRANS TF280, WirelessHART
6
1)
Preassembled Pt100: Class A (maximum MES: 0.15 + 0.002*|t| °C)
2)
Error calculation:
• Probable overall error = Ö(MES2 + AET2 + LTD2 + ATE2)
• Maximum overall error = MES + AET + LTD + ATE
• |t|: Absolute total of the measured temperature
• MES: Measuring error of the sensor
• AET: Measuring error of the transmitter (A/D conversion)
• LTD: Long-term drift
• ATE: Drift via ambient temperature
Displays and controls
Display (with illumination)
•Size of display 104 x 80 pixels
Number of digits
adjustable
Number of spaces after comma
adjustable
Setting options on site with 3 buttons
• with SIMATIC PDM
or HART Communicator
Auxiliary power
Battery 3.6 V DC
Communication
Wireless standard WirelessHART V7.1 conforming
Transmission frequency band
2.4 GHz (ISM-Band)
Range under reference conditions)
Up to 250 m (line of sight)
in outside areas
Up to 50 m (greatly dependent on
obstacles) in inside areas
Communication interfaces
• HART communication with
HART modem
WirelessHART
Certificates and approvals
Wireless communication approvals R&TTE, FCC
Pressure equipment directive This device does not fall under the
pressure equipment directive.
SITRANS TF280 WirelessHART temperature transmitter
Ordering data
Article No.
1)
Please order sensor separately.
Configuration
SITRANS TF280 WirelessHART
Temperature transmitter
7MP1110 - 0 A 777 - 0 7 7 0
(Required battery not included with
delivery, see accessories)
Connections/cable entry
Cable gland M20x1.5
1)
C
Sensor pipe with Pt100, G½"
male thread,
pre-mounted and connected
D
Display
Digital display, visible 1
Enclosure
Die-cast aluminum 1
Explosion protection
Not included A
Antenna
Variable, attached to device A
Further designs
Please add "-Z" to Article No.
and specify Order code(s) and
plain text.
Order code
Measuring point number (TAG No.)
max. 16 digits entered
in plain text
Y15: ....................
Y15
• Measuring point message
max. 27 characters entered in
plain text:
Y16: ....................
Y16
Accessories
Lithium battery
for SITRANS TF280/P280
7MP1990-0AA00
Mounting bracket, steel
7MF4997-1AC
Mounting bracket, stainless steel
7MF4997-1AJ
Cover, die-cast aluminum,
without window
7MF4997-1BB
Cover, die-cast aluminum,
with window
7MF4997-1BE
Thread adapter M20x1.5
(male thread) on ½-14 NP
(female thread)
7MP1990-0BA00
Thread adapter M20x1.5
(male thread) on G½B
(female thread)
7MP1990-0BB00
IE/WSN-PA LINK
with integral,
non-detachable antenna
6GK1411-6CA40-0AA0
IE/WSN-PA LINK
N connector for connection of
external antennas
6GK1411-6CA40-0BA0
HART modem with USB interface
7MF4997-1DB