UM344-2 MODEL DESIGNATION AND SPECIFICATIONS
September 1995
8
-
9
Power Supply
Minimum Compliance Voltage ................................+12 Vdc with no loop resistance, see Figure 4-4
Network (Current Sense or Loop)
Resistance (R
L
)................................................250 minimum, 1100 maximum
*R
L
= 43.4V
S
- 520.8; where V
S
= power supply
voltage
Maximum Loop Voltage .........................................+42 Vdc; for Ex N use, see Section 8.3.6
Ripple ...................................................................0.2 Vp-p, 47-125 Hz
Noise ....................................................................0.6 mV RMS maximum
Impedance ............................................................10 Ohms maximum
Power Supply Effect.....................................................Less than 0.005% of output span per volt
Indication, optional .......................................................4-1/2 digit LCD with status indicators
Transmitter Input Capacitance ......................................27,000 pF, HART C
n
= 6; refer to section 4.3.5
Network Topology
Point-To-Point
Transmitter Quantity ........................................1
Network Signal and Connection ........................Analog 4-20 mA, single current loop; see Figures
4-1 and 4-2
Network Resistance .........................................See Figure 4-4
Multi-Drop
Transmitter Quantity ........................................1 to 15
Network Signal and Connection ........................Digital, parallel connected; see Figure 4-3
Network Resistance .........................................See Figure 4-4
Transmitter I/O Isolation
Between Input/Output Terminal...............................250 Vac
Between Case and Networking Wiring.....................250 Vac
8.3.3 Two-Wire Cable
Type ...........................................................................Twisted Single-Pair, Shielded, Copper
Conductor Size for Network Length
Less than 5000 ft. (1500 m) ..............................24 AWG minimum
More than 5000 ft. (1500 m) .............................20 AWG minimum, 16 AWG maximum
Cable Capacitance .......................................................Refer to Section 4.3.5
Recommendation .........................................................Belden 8641, 24 AWG
Belden 8762, 20 AWG
Length, Maximum ........................................................Refer to Section 4.3.5