CALIBRATION AND MAINTENANCE UM344-2
September 1995
6
-
10
2) Disconnect the test resistance decade box and connect the test millivolt source to the transmitter.
See Figure 6-4.
3) Turn ON the power supply and proceed to section 6.1.4.2 and perform the Narrow/Wide Millivolt
calibration procedures.
6.1.4 Thermocouple/Millivolt Type Input Calibration
The calibration check consists of substituting a millivolt source for a thermocouple (TC) or mV (slidewire
potentiometer) type sensor. Millivolt values corresponding to Zero and Full Scale temperatures are applied
to the transmitter and the indicated output checked for accuracy.
Narrow/Wide Millivolt calibration consists of applying one low and one high millivolt signal to the
transmitter and entering them into EPROM.
Narrow Millivolt: Applicable to mV inputs between -11.00 to +26.00 millivolts including inputs from an
R, S, T, or B type thermocouple.
Wide Millivolt: Applicable to mV inputs between -18.00 to +103.00 millivolts including inputs from a
J, K, E, or N type thermocouple.
6.1.4.1 Thermocouple/Millivolt Calibration Check
1. Connect test equipment to transmitter as shown in Figure 6-4.
FIGURE 6-4 TC/mV Input Accuracy Check and Calibration Set Up
Bench Power
Supply (DC)
_
+
Millivolt
Source
+
_
250
MXC
+
_
+
_
Signal
Test
X02827S1 2 of 2
1
2 3
Model 344
Terminals
DVM
+
_