UM344-2 CALIBRATION AND MAINTENANCE
September 1995
Ohm input (wide range) ...... Equal to or less than +/-1.3 ohm of Zero and Full Scale ohmic
values.
Ohm input (narrow range) . Equal to or less than +/-0.45 ohm of Zero and Full Scale ohmic
values.
Equal to or less than +/-0.02% of selected span in ohms; the higher
value will be the accuracy tolerance.
5. Set decade box to resistance corresponding to Full Scale value and check accuracy of indication
against specifications listed in step 4.
If calibration is required, retain the test set-up and proceed to Section 6.1.3.2 to perform the Current
Source Calibration.
If accuracy is within specifications, calibration is not required. Disconnect test equipment, reconnect
all wires, and install enclosure cap.
6.1.3.2 Current Source Calibration (Transmitters with Software Rev 1 only)
1. Turn OFF the power supply.
Disconnect the test resistance decade box from the transmitter.
Connect the test microammeter to the transmitter as shown in Figure 6-3 and turn ON the power
supply.
2. From the MXC, execute the FIND XMTR program. (Refer to Section 2.)
Select CAL/TEST (F2) from the display menu.
1) Step through the MXC screens using the PREV and NEXT buttons to select CURRENT
SOURCE calibration. Press SELECT (F4). The MXC
will display the adjacent screen:
2) Press CONT (F4). MXC will display:
3) Press CONT (F4). MXC will display:
SHORT mV+ (1) TO RTD (2)
WITH COPPER WIRE
CONT
PRECISELY MEASURE THE
CURRENT BETWEEN RTDI
(2) AND mV- (3)
CONT
ENTER THE CURRENT
UA
QUIT ENTER
<