Page 80 SITRANS LU01 – INSTRUCTION MANUAL 7ML19985BE02
Parameters
P851 Short Shot Floor
Enter the minimum echo strength (in dB above 1 μV rms) derived from a short shot to be
considered for evaluation.
P852 Short Shot Range
Enter the maximum distance in Units (P005) to be measured using short shot echoes.
This feature is automatically altered when Transducer (P004) is altered.
Test (P90 0 to P913)
P900 Software Revision Number
View the EPROM Rev. #.
P901 Memory
Press ENTER to activate the SITRANS LU01 memory test.
Values
Range: 30 to 100
Preset: 50
Related
• P840 Short Shot Number
• P842 Short Shot Frequency
• P844 Short Shot Width
• P850 Short Shot Bias
• P852 Short Shot Range
Values Range: 0.000 to 9999
Altered By •P004 Transducer
Related
• P840 Short Shot Number
• P842 Short Shot Frequency
• P844 Short Shot Width
• P850 Short Shot Bias
• P851 Short Shot Floor
Note: These parameters are for Siemens Milltronics service personnel only.
Values Range: 00.00 to 99.99 (view only)
Values
Display: view only
PASS (memory test successful)
F1 RAM
F2 NOVRAM
F3 EEPROM
F4 EPROM