Vista de página 94
UMSITRPDS3-1 Post-Installation Test
November 2010
5-9
Transmitter above the active pressure taps (normal
arrangement)
Transmitter below the active pressure taps
(exception)
1
2
3
4
5
Transmitter
Compensation valve
Active pressure valves
Active pressure lines
Shutoff valves
7
9
10
Blowout valve
Condensation vessels (optional)
Active pressure taps
FIGURE 5-4 Measuring Gases
5.5.2.2 Measuring Liquids
Operate the shutoff valves in the following order. Refer to Figure 5-5 to identify the valves used in the
procedure.
1. Close all valves.
2. Open both shutoff valves (5) at the pressure taps.
3. Perform one of the following depending upon transmitter location:
“Transmitter below the active pressure tap” - Open both blowout valves (7) one after the other.
“Transmitter above the active pressure tap” - Open both vent valves (8) slightly until air-free
liquid emerges.
4. Close both blowout valves (7) or vent valves (8).
5. Open the active pressure valve (3A) and the vent valve at the high-pressure chamber of the
transmitter (1) slightly until air-free liquid emerges.
6. Close the vent valve.
7. Open the vent valve at the low-pressure chamber of the transmitter (1) slightly until air-free liquid
emerges.