LIQUID APPLICATIONS - VOLUME
1. Beam should not detect bin bottom. If this occurs use range extension parameters
(on transceivers where available) to omit false echoes. A 6° beam angle represents a
rise : run of about 20 : 1 (10 : 1 for 12°). In most tanks, the transducer should be
centered as much as possible (without interference from inlet) for optimum reading range.
2. Sound beam must be perpendicular to liquid surface.
If standpipe is used, refer to Liquid / Applications - Standpipes.
3. Echo has missed improperly leveled transducer.
4. When performing an empty or full calibration, the tank must contain its normal vapour and
be at its normal temperature.
tank manufacturer’s
empty level
discharge
Empty level for
‘alternate’ location.
Below this level, echo
would reflect away
from the transducer.
may require target to
obtain empty reading
rise
span : corresponds to
tank manufacturer’s
empty level.
beam
angle
tank
manufacturer’s
full level
Maintain full level
for full calibration.
Above this level
erroneous readings
will result as level
has entered
blanking zone.
(shaded area)
‘ Alternate ’
‘ Preferred ’
‘ Bad ’
7ML19981AK01.2 XPS/XCT Transducers 15